Covering all the existing test methods for the filter integrity testing High-definition, large-size color touch screen input more abundant test information, and can automatically select and match suitable for the current filter detection program Digital sensor makes the test results more accuracy and consistency The CPU of the instrument, adopts industrial grade dual core high speed and high stability ensure the safety, reliability and rapidity of the instrument in the use process, and solve the problems of instability and system vulnerability caused by the operating system “pre-program” improves the integrity of the test parameters, greatly simplifies the user’s operation procedures, and truly realizes “one key completion” With detailed test data and complete test curve, users can comprehensively and accurately analyze a variety of performance indicators of the filter under test, and provide objective analysis report for users Friendly Interface and prompt information make the operation simple and convenient, A variety of input methods: Chinese, English, numbers and punctuation Scientific electronic signature and user management for division of responsibilities and prevent misoperation With digital and analog interface (RS232/USB), also customize a variety of industrial bus and simulation control port according to customer’s need Carrying out the design of specific solutions according to the needs of customers; Many years of experiences and strong professional services provide the technical support for the design and configuration of the filter system for customers

Power Supply100–260 VAC, 50HZ, 120W
Operation Pressure100-10000 mbar (145psi)
UnitMbar
Operation ConditionAmbient Temperature:+5℃ ~ +40℃;Relative Humidity:10-80%
Dimension(mm)400(Depth) ×260(Width) × 240(Height1)x220(Height2)
FunctionManual Bubble point Test;Basic Bubble Point Test;Extensive Bubble Point Test; Pressure Holding Test;Diffusion FLow Test;Water Immersion Test;Ultrafitration Membrane Test
Test AccuracyNet Volume Test:≤ ± 4%;Bubble Point Test:≤ ± 50mbar ; Diffusion Flow Test:≤ ± 4% ;Water Immersion:≤ ± 0.01ml
Test ScopeBP:100-8000mbar   DF:1-600ml/min  WI:0.01-100ml/min
Applied RangeSymmetric and asymmetric membrane test, syrine filters, capsule filters, plate filters, cartridge filters ;ultrafiltration membrane packages, ultrafiltration columns, various irregular filters
Audit TrialAudit trail records can exportable and ireeversible
Authority Management4 level User Name and Password to Login ,electronic signature to meet the FDA 21 CFR PART 11 requirements
Reservation Solution(programs)120sets

Audit Trial

(customization)

More subject logs
Print FunctionBuilt-in printer
Applicable EnvironmentAbove D level
Weight8KG
Historical RecordMore than 5000 sets unlimited sets records
Records CopySupport USB disk to export (include test curve);
Display ScreenHigh definition 7 inch ,color touch screen;
Serial ConnectionRS232/USB
LanguageChinese/English
Work ModeOnline/Offline
Dust and splash levelIP54
Signal output(4-20)mA, RS485,12V Alarm output

 